So, You Just Bought an Oscilloscope: Your First Measurements

By TAM Editorial Staff ·

In this article
  1. First, the important safety rule
  2. What all those parts actually do
  3. Understand the probe before connecting it
  4. Measurement one: display a square wave
  5. Measurement two: read voltage, period and frequency
  6. Measurement three: use cursors for your own measurement
  7. Measurement four: look at an audio signal
  8. When the display looks wrong
  9. What the specifications mean, briefly
  10. Where to go next
  11. References

You bought an oscilloscope. The screen lights up, there are enough knobs to suggest an aircraft cockpit, and the manual is hundreds of pages long. Fortunately, the essentials are simpler than the front panel suggests. An oscilloscope displays voltage against time. Once you understand vertical scale, horizontal scale and triggering, you can make useful measurements almost immediately.

We will use the Siglent SDS1202X-E as our example. It is a two-channel, 200 MHz digital storage oscilloscope with a 7-inch display, a maximum real-time sample rate of 1 GSa/s and up to 14 Mpoints of waveform memory. Those figures matter, but they are not prerequisites for taking your first measurement. The techniques apply to most conventional bench oscilloscopes.

Editor’s photo: Full view of the SDS1202X-E on your bench, powered on, with the principal control groups visible.

First, the important safety rule

The ground clip on an ordinary bench oscilloscope probe is not a floating measurement lead. On the SDS1202X-E it is referenced to the instrument’s protective earth. Connecting it to the wrong point can short part of a circuit to earth, damage equipment and create a shock or fire hazard.

  • For these exercises, use only the scope’s built-in probe-compensation output and, later, a low-voltage, isolated signal source.
  • Do not clip the probe ground to the hot side of a wall outlet, the primary side of a power supply, or an unidentified point in mains-powered equipment.
  • Do not defeat the scope’s safety-earth connection or use a two-prong adapter to make it “float.”
  • Two ordinary probe ground clips on this scope share ground; the channels are not independently isolated.
  • For mains, bridge-tied amplifier outputs, floating supplies and other hazardous measurements, use an appropriate rated differential probe or isolation-rated instrument and follow its procedures. Do not improvise.

What all those parts actually do

The display and its graticule

The screen is divided into a grid called the graticule. The horizontal axis represents time; the vertical axis represents voltage. A waveform two major divisions high at 500 mV/div represents 1 V between those two levels. A cycle spanning four divisions at 250 µs/div lasts 1 ms, corresponding to 1 kHz.

The screen also shows the channel’s volts-per-division setting, the timebase, probe attenuation, trigger status and other information. Learn to read those labels before interpreting a trace.

CH1 and CH2: the signal inputs

The two BNC connectors accept the probe leads. Each channel has its own vertical controls, so you can display and compare two low-voltage signals at once. The separate external trigger connector is for triggering, not a third measurement channel.

Vertical controls: how tall is the waveform?

  • CH1 / CH2 buttons: enable a channel and open its settings, including coupling and probe attenuation.
  • Scale (volts/div): sets how much voltage one vertical grid division represents. A smaller value makes the same waveform appear taller.
  • Position: moves the trace up or down without changing the underlying signal.
  • DC coupling: shows both the DC level and any AC variation. This is the default starting point for most measurements.
  • AC coupling: blocks the DC component so a small AC signal riding on a large offset is easier to examine. It changes what you are measuring and attenuates sufficiently low frequencies.

Horizontal controls: how much time are you viewing?

  • Timebase (seconds/div): determines how much time each horizontal division represents. At 1 ms/div, ten divisions cover 10 ms.
  • Horizontal position: shifts which part of the acquired record appears on the screen, including data before or after a trigger.
  • Zoom and memory: once you’re comfortable with the basics, these let you inspect smaller details in a long recording.

Trigger controls: what keeps the trace still?

A scope repeatedly captures signals. The trigger establishes a consistent reference point so a periodic waveform stays in place instead of drifting sideways.

  • Source: choose the channel containing the signal, usually CH1 for your first test.
  • Type: start with an edge trigger.
  • Slope: rising means the trigger occurs as voltage crosses the selected level going up; falling means going down.
  • Level: set a voltage the signal actually crosses, often near its midpoint.
  • Auto / Normal / Single: Auto generally keeps updating even without a valid trigger; Normal waits for a valid trigger; Single captures one event and then stops.

Don’t confuse trigger coupling with channel coupling. The former affects the circuit used to decide when to capture. The latter affects the waveform displayed on a channel.

The other controls worth locating

  • Auto Setup: a useful shortcut for finding a signal. It is not proof that every resulting setting is appropriate.
  • Run/Stop: freezes or resumes acquisitions so you can inspect a captured waveform.
  • Measure: gives numerical readings such as frequency, period, peak-to-peak voltage, maximum, minimum and RMS voltage.
  • Cursors: manually place markers to measure differences in voltage or time.
  • Universal knob and menu softkeys: select and change the options displayed in menus.
  • USB port: convenient for saving screenshots or waveform data; follow the instrument’s Save/Recall menu instructions.
  • Probe-compensation output and ground terminal: the built-in safe test signal we will use next.

Editor’s photo: Close-up of the SDS1202X-E front panel, with the vertical, horizontal and trigger groups labeled in the image or caption.

Understand the probe before connecting it

A passive 10× probe does not simply act as a piece of wire. It attenuates the signal by a factor of ten and, when properly configured, reduces the loading of many circuits compared with a 1× setting. The scope must know whether the probe is set to 1× or 10× so the displayed voltage is correct.

  • Set the probe’s physical switch to 10×.
  • Set the selected scope channel’s probe attenuation to 10× as well. A mismatch can make the voltage reading wrong by a factor of ten.
  • Connect the probe’s BNC connector to CH1, push it into place and twist the bayonet collar to lock it.
  • Inspect the probe tip, insulation and ground lead for damage before use.

Measurement one: display a square wave

The scope supplies a calibration-style square wave specifically for checking probes. This is the ideal first experiment because you do not need to open another piece of electronic equipment.

  1. Turn on the SDS1202X-E and allow it to finish starting up. If its settings are unfamiliar, use the Default control to reset to a known starting point.
  2. With the probe attached to CH1 and set to 10×, connect its ground clip to the front-panel ground terminal beside the probe-compensation output.
  3. Touch or clip the probe tip to the probe-compensation signal output. Do not connect it to another connector by guesswork.
  4. Press Auto Setup. A repetitive square wave should appear. If necessary, confirm that CH1 is enabled and its probe factor matches the probe.
  5. Adjust Volts/div until the trace occupies several vertical divisions; adjust Time/div until several complete cycles are visible.
  6. Check the corners and the flat top of the square wave. An undershoot or rounded top can indicate an undercompensated probe; overshoot can indicate overcompensation. Follow the probe’s instructions and adjust its small compensation trimmer with a nonmetallic tool until the waveform is properly squared.
  7. Repeat with the second probe on CH2, one at a time, if you intend to use both channels.

Editor’s photos: (1) Probe tip and ground clip connected to the compensation terminals; (2) undercompensated, correctly compensated and overcompensated square waves, taken from your actual scope.

Measurement two: read voltage, period and frequency

Leave the probe on the compensation output. There are two useful ways to extract measurements.

Method A: count divisions

  1. Note the vertical scale shown on the screen.
  2. Count the vertical divisions between the square wave’s low and high levels, then multiply by volts/div. That is the approximate peak-to-peak voltage (Vpp).
  3. Note the timebase and count the horizontal divisions for one complete cycle. Multiply by seconds/div to obtain the period (T).
  4. Calculate frequency from f = 1/T. A 1 ms period corresponds to 1,000 Hz, or 1 kHz.

For example, if a waveform spans three divisions at 1 V/div, it is approximately 3 Vpp. If its cycle spans four divisions at 250 µs/div, the period is 1 ms and the frequency is 1 kHz. These are illustrative numbers: use the actual values shown on your own instrument rather than assuming its compensation output is calibrated to these exact settings.

Method B: let the scope measure it

  1. Open Measure.
  2. Select CH1 as the measurement source.
  3. Add Vpp, Frequency and Period, using the on-screen menu and universal knob as needed.
  4. Compare the automatic results with your graticule calculation. Small differences are normal because counting grid divisions by eye is approximate.

Editor’s photo: A properly compensated square wave with Vpp, frequency and period displayed in the measurement readouts.

Measurement three: use cursors for your own measurement

Automatic measurements are convenient, but they cannot read your mind. Cursors let you choose exactly which features to compare.

  1. Press Cursors and turn cursor mode on.
  2. Select CH1 as the source.
  3. Choose the vertical time markers (X cursors). Place one at a rising edge and the other at the next rising edge. The displayed difference, Δt, is one cycle’s period.
  4. Switch to the horizontal voltage markers (Y cursors). Put one on the waveform’s low plateau and the other on the high plateau. The difference, ΔV, is its approximate peak-to-peak voltage.

Editor’s photo: One screenshot showing time cursors on adjacent rising edges, and another showing voltage cursors on the square wave’s high and low levels.

Measurement four: look at an audio signal

Now try a signal that is relevant to a recording studio: a 1 kHz sine wave from an isolated, low-voltage audio signal generator or an appropriately interfaced line-level output. Start with a conservative amplitude, such as approximately 1 Vpp, and verify the source’s ground reference before connecting anything. If you aren’t certain the signal and its return are safe to connect to earth, stop here.

  1. Connect the probe ground to the source’s verified signal ground, and the probe tip to its signal output. Keep the scope’s earth connection intact.
  2. Start on DC coupling, with CH1 at a suitable scale such as 500 mV/div and the timebase around 200–500 µs/div.
  3. Use Auto Setup if necessary, then adjust the scales manually so you can see two or three complete cycles.
  4. Choose an edge trigger on CH1 and set the trigger level near the center of the sine wave to stabilize the display.
  5. Enable frequency and RMS voltage measurements. A clean sine wave is smooth and symmetrical; significant clipping will flatten its peaks.
  6. As an exercise, gradually increase the source level within safe limits and watch the displayed amplitude change.

Be careful with RMS readings: a sine wave has Vrms = Vpeak/√2, or approximately Vpp/(2√2), when measured without DC offset. For a 2 Vpp sine wave, that is about 0.707 Vrms. The relationship does not apply to a square wave or a distorted waveform in the same way.

Editor’s photos: Clean 1 kHz sine wave from a line-level source; optionally a separate clipping example, clearly identified as deliberately generated and safely low-voltage.

When the display looks wrong

  • Nothing on screen: check the probe connection, channel enable, volts/div setting, probe attenuation and whether the scope is stopped.
  • Trace rolling sideways: check the trigger source, slope and level. Auto Setup is useful for recovering a stable starting point.
  • Voltage ten times too high or low: compare the physical probe switch with the channel’s attenuation setting.
  • Square wave with rounded or peaked corners: check probe compensation, connections and bandwidth/loading before blaming the signal source.
  • Unexpected offset: verify whether the channel uses DC or AC coupling and where the ground reference is connected.
  • Very noisy or distorted trace: shorten the probe ground connection where practical, verify the source, and consider bandwidth limits and sample rate. A scope can reveal noise introduced by a poor probing arrangement.
  • Frequency measurement jumping around: capture more than one complete cycle, adjust the trigger and inspect whether the waveform is actually periodic.

What the specifications mean, briefly

  • 200 MHz bandwidth describes the scope’s analog frequency response, not the highest frequency you can measure with complete fidelity. Fast edges contain harmonics far above their repetition frequency.
  • 1 GSa/s sampling means up to one billion samples per second under supported configurations; actual sample rate depends on setup and acquisition settings.
  • 14 Mpoints memory makes it possible to capture long records while retaining useful time detail, within the instrument’s acquisition limits.
  • Two channels let you compare input and output, timing and phase relationships, or two points in a circuit—provided both ground references can be connected safely.

Where to go next

Once you can capture the built-in square wave, measure a period, read Vpp and stabilize an audio sine wave, you already know the controls you will use most often. Next experiments might include comparing input and output levels on a low-voltage audio circuit, identifying clipping, observing attack and decay of a control signal, or using FFT mode to inspect a signal’s spectrum. Start with predictable signals and a sound understanding of the ground connection. A scope is most useful when you know what question you are asking it.

References